Cell Booster

General Details

“Cell Booster” is SMPS based with isolated output, battery charger with settable charging voltage and current control feature. It can be used for various battery current and voltage ratings. The charging voltage and current can be set from 2V to 18V, 4A to 50A. Other ratings of voltage and current can be made available on demand. The Cell Booster is a microcontroller based system. Calibration parameters provided are password protected. Users can set battery charging voltage and current and it monitors all parameters. The unit is well protected from short circuit and battery reverse polarity

Product Highlights

  • Front end PFC for high power factor (near unity).
  • Low input ATHD, <20%.
  • Low output voltage ripple.
  • Input voltage range 180 – 265V.
  • Constant power operation.
  • 20×4 character LCD display.
  • Programmable output current & voltage.
  • Operation as voltage source or current source..
  • Calibration of measurements from front panel.
  • Short circuit protected.
  • Soft start.
  • Battery reverse protection is in-built.

Technical Info
Model Number Output
voltage
in volts
Output current, in Amps 0/p power max, in watts Input AC Operating Voltage Range Power factor Input
ATHD
Operation
mode
Efficiency Casing Box dimensions in ‘mm’           Protections
SCB14V20A 2-14V 2-20A 280 180 – 280V >0.95 <15% CV, CC >85% Metal 295X434X120 0/P Short ckt,0/P overvoltage Input overvoltage, No load
SCB14V
40A
2-14V 4-40A 560 180-280V >0.95 <15% CV,CC >85% Metal 295X434X120 0/P Short ckt,0/P over voltage,Input overvoltage,
No load
SCB18V
25A
2-18V 2-25A 450 180-280V >0.95 <15% CV,CC >85% Metal 295X434X150 0/P Short ckt,0/P overvoltage,Input over voltage,No load
SCB18V
50A
2-18V 2-50A 900 180-280V >0.95 <15% CV,CC >85% Metal 295X434X150 0/P Short ckt,
0/P over voltage,
Input over voltage,
No load
Applications
  • As battery chargers in Railways, Industrial Applications